INDICATIONS FOR <strong>CPAP</strong>:<br />

• Acute respiratory failure<br />

• Pulmonary contusion and flail chest<br />

• Cardiogenic pulmonary oedema<br />

• Asthma<br />

• Chronic airways disease<br />

• Post-operative atelectasis<br />

• Obstructive sleep apnoea<br />

• Pneumonia<br />

CONTRAINDICATIONS:<br />

• Unstable facial fractures<br />

• Extensive facial lacerations<br />

• Laryngeal trauma<br />

• Recent tracheal or oesophageal<br />

anastomosis<br />

• Basal skull fracture, (use with caution)<br />

• Patients at risk of GIT bleeds/ileus<br />

USE WITH CAUTION<br />

• Patients with recent GI surgery<br />

• Decreased level of consciousness<br />

EQUIPMENT:<br />

Drager <strong>CPAP</strong> machine/Drager Ventilator with NIV function<br />

Ventilation tubing for wet circuit (humidification base in tubing pack)<br />

Sterile water bag<br />

<strong>CPAP</strong> mask and head strap<br />

PEEP valve (5,7.5, 10 or 12.5, may start with 10cmH20)<br />

Bacterial/viral filter (machine side of tubing)<br />

Test lung<br />

White rubber connector (for dedicated Drager <strong>CPAP</strong> unit only)<br />

<strong>CPAP</strong> MACHINE<br />

SET UP:<br />

See pictures below for assistance<br />

• Keep “Y” connector attached to ventilator tubing<br />

• Remove white side of vent tubing<br />

• Place bacterial/viral filter to <strong>CPAP</strong> machine, then attach short arm of blue tubing from filter<br />

to humidification bath<br />

• Attach long arm of blue tubing from humidification bath to <strong>CPAP</strong> mask<br />

• Assemble blue wire of humidification bath holder to blue tubing<br />

• Using white rubber connector attach PEEP valve to exposed end of “Y” connector<br />

• Perform safety valve check (see below)<br />

• Adjust prescribed Fi02 (use flow rate of 40L/min)<br />

Equipment for Designated <strong>CPAP</strong> circuit<br />

©ROYAL PRINCE ALFRED HOSPITAL INTENSIVE CARE SERVICE