Human Med AG<br />

<strong>Instructions</strong> <strong>for</strong> <strong>use</strong> and processing of the LipoCollector TM<br />

The parts are now as shown below:<br />

Page:<br />

Vers./Date:<br />

8. The fine filters (A) are single-<strong>use</strong> material and must be disposed of.<br />

8 of 20<br />

1 / 15.07.2008<br />

9. All individual parts required <strong>for</strong> re-<strong>use</strong>, collection container, per<strong>for</strong>ated metal<br />

plate, container unit, lid, reducer and retaining ring, must be immersed in the<br />

cleaning and disinfecting solution (Gigasept Instru AF, 1.5%, made up with lukewarm<br />

water, demineralised water if possible) <strong>for</strong> at least 10 minutes or longer if necessary<br />

<strong>for</strong> stubborn or dried soiling.<br />

A<br />

10. Then the cavities must be cleaned out:<br />

Both tubes of the container<br />

insert must be cleaned<br />

with the 5 mm brush.