AquaShape LipoCollector™ Instructions for use
AquaShape LipoCollector™ Instructions for use
AquaShape LipoCollector™ Instructions for use
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Human Med AG<br />
<strong>Instructions</strong> <strong>for</strong> <strong>use</strong> and processing of the LipoCollector TM<br />
The parts are now as shown below:<br />
Page:<br />
Vers./Date:<br />
8. The fine filters (A) are single-<strong>use</strong> material and must be disposed of.<br />
8 of 20<br />
1 / 15.07.2008<br />
9. All individual parts required <strong>for</strong> re-<strong>use</strong>, collection container, per<strong>for</strong>ated metal<br />
plate, container unit, lid, reducer and retaining ring, must be immersed in the<br />
cleaning and disinfecting solution (Gigasept Instru AF, 1.5%, made up with lukewarm<br />
water, demineralised water if possible) <strong>for</strong> at least 10 minutes or longer if necessary<br />
<strong>for</strong> stubborn or dried soiling.<br />
A<br />
10. Then the cavities must be cleaned out:<br />
Both tubes of the container<br />
insert must be cleaned<br />
with the 5 mm brush.