AquaShape LipoCollector™ Instructions for use
AquaShape LipoCollector™ Instructions for use
AquaShape LipoCollector™ Instructions for use
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Human Med AG<br />
<strong>Instructions</strong> <strong>for</strong> <strong>use</strong> and processing of the LipoCollector TM<br />
Processing and assembly of the LipoCollector TM<br />
Special Note:<br />
ATTENTION!<br />
Page:<br />
Vers./Date:<br />
5 of 20<br />
1 / 15.07.2008<br />
Plastics can be damaged by highly alkaline (pH > 10) cleaning and disinfecting<br />
agents.<br />
Only those cleaning agents should be <strong>use</strong>d which are recommended <strong>for</strong> the cleaning<br />
of plastics. Please refer to the manufacturer of the instrument cleaning or disinfecting<br />
agent.<br />
Do not <strong>use</strong> cleaning or disinfection agents that contain phenol.<br />
We recommend the <strong>use</strong> of a combined cleaning and disinfecting agent.<br />
Agents tested by us are listed on page 10.<br />
Validation was carried out with Gigasept Instru AF.<br />
Cleaning and sterilisation must be carried out only by trained personnel in designated<br />
rooms.<br />
Limit to re -<strong>use</strong>: The end of the product lifetime is usually determined by wear and tear resulting from<br />
<strong>use</strong>.<br />
All components of the LipoCollector TM have safely tested <strong>for</strong> 50 processing cycles.<br />
Any additional processing cycles become the responsibility of the <strong>use</strong>r.<br />
Necessary tools Ultrasound cleaning bath, rinsing bath, cleaning bath, cleaning brushes as follows<br />
<strong>Instructions</strong>:<br />
At location of <strong>use</strong>: Remove surface soiling with a disposable cloth / paper towel.<br />
Storage and transport: It is recommended to prepare the parts of the containers <strong>for</strong> the next <strong>use</strong> immediately<br />
after <strong>use</strong>.<br />
Dismantling and<br />
pre-cleaning:<br />
1.<br />
10 mm<br />
5 mm<br />
Remove the reducer (A) on the “PATIENT” connection from the lid of the container.<br />
A