Human Med AG<br />

<strong>Instructions</strong> <strong>for</strong> <strong>use</strong> and processing of the LipoCollector TM<br />

Processing and assembly of the LipoCollector TM<br />

Special Note:<br />

ATTENTION!<br />

Page:<br />

Vers./Date:<br />

5 of 20<br />

1 / 15.07.2008<br />

Plastics can be damaged by highly alkaline (pH > 10) cleaning and disinfecting<br />

agents.<br />

Only those cleaning agents should be <strong>use</strong>d which are recommended <strong>for</strong> the cleaning<br />

of plastics. Please refer to the manufacturer of the instrument cleaning or disinfecting<br />

agent.<br />

Do not <strong>use</strong> cleaning or disinfection agents that contain phenol.<br />

We recommend the <strong>use</strong> of a combined cleaning and disinfecting agent.<br />

Agents tested by us are listed on page 10.<br />

Validation was carried out with Gigasept Instru AF.<br />

Cleaning and sterilisation must be carried out only by trained personnel in designated<br />

rooms.<br />

Limit to re -<strong>use</strong>: The end of the product lifetime is usually determined by wear and tear resulting from<br />

<strong>use</strong>.<br />

All components of the LipoCollector TM have safely tested <strong>for</strong> 50 processing cycles.<br />

Any additional processing cycles become the responsibility of the <strong>use</strong>r.<br />

Necessary tools Ultrasound cleaning bath, rinsing bath, cleaning bath, cleaning brushes as follows<br />

<strong>Instructions</strong>:<br />

At location of <strong>use</strong>: Remove surface soiling with a disposable cloth / paper towel.<br />

Storage and transport: It is recommended to prepare the parts of the containers <strong>for</strong> the next <strong>use</strong> immediately<br />

after <strong>use</strong>.<br />

Dismantling and<br />

pre-cleaning:<br />

1.<br />

10 mm<br />

5 mm<br />

Remove the reducer (A) on the “PATIENT” connection from the lid of the container.<br />

A