<strong>Instructions</strong> <strong>for</strong> <strong>use</strong> and processing of the LipoCollector TM<br />

Page:<br />

Vers./Date:<br />

10 of 20<br />

1 / 15.07.2008<br />

12. All other surfaces must then be cleaned with the 10 mm brush.<br />

Pre-cleaning with the brushes must be carried out using the cleaning and disinfecting<br />

solution (Gigasept Instru AF, 1.5%, made up with lukewarm water, demineralised<br />

water if possible).<br />

These steps must be repeated until the brushes and the surfaces and cavities to be<br />

cleaned are free of visible contamination.<br />

The instrument cleaning solution must be renewed at least daily or if there is visible<br />

contamination.<br />

13. Cleaning and<br />

disinfection:<br />

ATTENTION!<br />

14. Drying:<br />

15. Maintenance:<br />

16. Packing:<br />

17. Sterilisation:<br />

18. Storage:<br />

All parts must now be completely immersed in an ultrasound bath with cleaning and<br />

disinfecting solution (Gigasept Instru AF, 3%, made up with lukewarm water,<br />

demineralised water if possible). The cleaning process in the ultrasound bath must be<br />

at least 5 minutes long.<br />

Cleaning <strong>for</strong> 15 minutes in an immersion bath is also possible. In this case the same<br />

cleaning solution and the same temperature are required.<br />

Do not mix with other cleaners.<br />

After cleaning in the solution the parts must be thoroughly rinsed with running water<br />

(preferably demineralised) or in a rinsing bath <strong>for</strong> at least 1 minute.<br />

Repeat rinse step again.<br />

Dry all parts inside and out with sterile compressed air.<br />

Maintenance of the individual parts is not intended.<br />

The dried individual parts of the LipoCollector TM must be packed as follows in<br />

suitable sterile packaging suitable <strong>for</strong> steam sterilisation as per ISO 11607. The<br />

package must be large enough that the seal is not under tension.<br />

1. Container insert separately<br />

2. Lid separately<br />

3. Retaining ring, per<strong>for</strong>ated metal plates and reducer together<br />

4. Collection container separately<br />

Steam sterilisation using a fractionated vacuum procedure at 121°C <strong>for</strong> at least 15<br />

minutes and at 134°C <strong>for</strong> 3 minutes is validated.<br />

Sterilisation at 134°C <strong>for</strong> at least 10 minutes was tested with no negative effects on<br />

the material detected.<br />

For every sterilisation check the pressure protocol <strong>for</strong> the particular steriliser to ensure<br />

compliance with the necessary parameters.<br />

The individual sterile foil bags must be stored in a closed cupboard protected from<br />

dust, moisture and large temperature fluctuations.<br />

The shelf life is determined by the specifications <strong>for</strong> the sterile packaging <strong>use</strong>d.