Dynamic Results<br />
Ion Detection Test<br />
Starting All Diagnostics Scan Tests<br />
16:19:31: Start scan readback test on device Trap Offset (V) -- -132 to 132<br />
16:19:36: Scan readback test ended<br />
16:19:38: Result: PASSED<br />
16:19:38: Start scan readback test on device Tube gate (V) -- -200 to 198.01<br />
16:19:43: Scan readback test ended<br />
16:19:45: Result: PASSED<br />
16:19:45: Start scan readback test on device Multiplier (V) -- 0 to -2200<br />
16:19:58: Scan readback test ended<br />
16:20:00: Result: PASSED