<strong>LCQ</strong> Recommended Maintenance Schedule<br />
Every 3 months:<br />
Change the mech. pump oil.<br />
6 months to a year:<br />
Change the turbo pump oil.<br />
Notes<br />
- The API stack and analyzer should be clean as needed.<br />
- If the sensitivity starts to drop off and can not be<br />
restored, clean the API stack (and analyzer) as needed.