Detection System Parameters<br />

◗ Dynode Voltage: 15kV (fixed)<br />

◗ Electron Multiplier: 2500 Volts max.<br />

Note:<br />

– Multiplier voltage is a calibrated parameter.<br />

– For best results, the multiplier should not be set manually.<br />

<strong>Troubleshooting</strong>:<br />

No peaks on Classic: Check multiplier voltage<br />

Switch dynode polarity<br />

Check for ions<br />

No peaks on <strong>Deca</strong>/<strong>Duo</strong>: Check Multiplier and Dynode Voltages.<br />

Noisy spectra with heated capillary capped off:<br />

Potential Dynode noise<br />

Clean Dynode Cup