Cal Mix Tune Parameters<br />

<strong>LCQ</strong> Ion Optics<br />

Octapole 1 Offset (V) for Classic/<strong>Duo</strong>: -1 to -5<br />

Octapole 1 Offset (V) for <strong>Deca</strong>: -4 to -9<br />

Lens Voltage (v): -16 to -50<br />

Octapole 2 Offset (V) for Classic / <strong>Duo</strong>: -5.5 to -10<br />

Octapole 2 Offset (V) for <strong>Deca</strong>: -7 to -15<br />

Octapole RF Amplitude (V p-p): 400<br />

Entrance Lens (V) for the <strong>Deca</strong> only: -35 to -60 V<br />

<strong>Troubleshooting</strong>:<br />

Loss of Sensitivity: * Clean octapoles(multipoles) and lens<br />

Octapole Diagnostic errors: * Tune multipole RF prior to running diagnostics.<br />

* Multipole RF tune now performed in calibration<br />

for all <strong>LCQ</strong>s run with Xcalibur.