<strong>Troubleshooting</strong> the API Stack<br />
◗ Loss of Sensitivity: Clean skimmer and tube lens<br />
◗ Low Convectron Gauge Pressure: Heated capillary plugged<br />
◗ Cal Mix Contamination: Clean API probe and heated capillary area<br />
◗ Spiky noise: Bent capillary tip or dirty heated capillary<br />
◗ Constant Background: Possible heated capillary contamination<br />
◗ Overall solution for 90% of API Stack related Problems:<br />
� Good maintenance:<br />
Caution:<br />
Clean/Rinse the heated capillary region<br />
and spray shield daily.<br />
� Use of the divert valve<br />
� Use of Spray Cap and Orthogonal<br />
Sampling Adapter<br />
Periodically empty the waste bottle to avoid potential back streaming of waste<br />
solvent into the source region.