API-2 API 2 / Off Axis ESI Guidelines<br />

LC Flow Rates<br />

Infusion or LC at flow<br />

rates of 200 µL/min<br />

Slide<br />

Plate<br />

Position<br />

Operational Parameters<br />

Probe<br />

Position<br />

(1 to 7)<br />

Heated<br />

Capillary<br />

Temperature<br />

Top 4 Typical setting:<br />

250 °C<br />

Top 4 Typical setting:<br />

350 °C<br />

Top 5 Typical setting:<br />

350 °C<br />

Sheath<br />

Gas<br />

Required<br />

Typical setting:<br />

10 to 30 units<br />

Required<br />

Typical setting:<br />

80+ units<br />

Required<br />

Typical setting:<br />

80+ units<br />

Auxiliary<br />

Gas<br />

Not required<br />

Typical<br />

setting:<br />

0 units<br />

Required<br />

Typical<br />

setting:<br />

10 to 20 units<br />

Required<br />

Typical<br />

setting:<br />

10 to 20 units