Cal Mix Tune Parameters<br />

API-2 on <strong>LCQ</strong> <strong>Deca</strong> / <strong>LCQ</strong> <strong>Duo</strong><br />

◗ Infusion Flow Rate (µL/min.): 3-5<br />

◗ Spray Voltage (kV): 4-6<br />

◗ Spray Current (µA): 0.1-0.75<br />

◗ Sheath Gas Flow Rate (arb): 20-40<br />

◗ Aux. Gas Flow Rate (arb): 0<br />

◗ Capillary Temperature: 200-250°C<br />

◗ Probe Position : 2