EN9.8 Accessing Saved Measuring DataYou can save up to 1024 measuring data directly in the instrument’s memory. If thememory contains 1024 saved values, further values will overwrite the oldest savedvalues. Each saved value is given a consecutive 4-digit number. Using this numberyou can see whether the instrument has overwritten old entries. This is the case if theconsecutive number is larger than 1024. If you delete all measuring data, the counterstarts again at 0001.To see how many values are saved you can select "Device Information" (seeChapter 11.3).1. Press and select "Measuring Data".The list of results is displayed (see Fig. 9 - 3).Fig. 9 - 3Measuring data50 C96IB01
9.9 Exporting Measuring Data to a PCENExporting measuring data to a PC does not require a special program. You just needa PC with an infrared interface or a PC with an IrDA USB adapter installed andconnected. Contact your Anton Paar representative if you need an IrDA USB adapterfor your PC.To set the data export format1. Press and select "Setup > Import / Export > Data Format".You can choose between different format options.2. Choose the data format, or , and press .3. Choose the delimiter, [comma], [semi-colon] or [slash], and press.4. Choose the decimal point, [point] or [comma], and press .Your changes are accepted and used for the following data export. The settingsfor delimiter and decimal point are only applied with CSV data exports.To export a measured value1. If your PC has no IrDA interface, install the IrDA USB adapter on your PC.2. Press and select "Measuring Data".3. Select the measured value required for export using the arrow keys.4. Press .5. Hold the IrDA interface of DMA 35 to the IrDA interface of the PC to establishcommunication between the two instruments.6. Select to export the selected measured value to the PC.A dialog window appears on the PC asking if you want to accept the file.7. Click .The measured value is transferred and filed on the desktop.To export all measuring dataYou have two options for exporting all measuring data to a PC.Option 1:1. If your PC has no IrDA interface, install the IrDA USB adapter on your PC.2. Press and select "Measuring Data".3. Press .4. Hold the IrDA interface of DMA 35 to the IrDA interface of the PC to establishcommunication between the two instruments.C96IB01 51
- Seite 1: DMA 35Portable Density/Specific Gra
- Seite 4 und 5: ENWhile every precaution has been t
- Seite 6 und 7: EN9.1.3 Importing Custom Functions
- Seite 8 und 9: EN1 About the Instruction ManualThi
- Seite 10 und 11: EN2 Safety Instructions• Read thi
- Seite 12 und 13: EN• Do not leave sample/rinsing c
- Seite 14 und 15: EN5 Checking the Supplied PartsDMA
- Seite 16 und 17: ENTable 5.2: Optional partsArticle
- Seite 18 und 19: EN2 314Fig. 6 - 2 Top view of DMA 3
- Seite 20 und 21: EN6.3 Display23b3a1412561171098Fig.
- Seite 22 und 23: EN7 Operating the DMA 357.1 Connect
- Seite 24 und 25: EN8 Defining the Basic Settings8.1
- Seite 26 und 27: EN8.7 Allocation of the Softkey Fun
- Seite 28 und 29: ENTo set the time format1. Press a
- Seite 30 und 31: EN• Make sure to carefully clean
- Seite 32 und 33: ENMeas. OptionsDescriptionunitsBaum
- Seite 34 und 35: EN9.2 Selecting Measuring MethodsAp
- Seite 36 und 37: ENMethod list templateThe following
- Seite 38 und 39: EN9.2.6 Selecting a Method1. Press
- Seite 40 und 41: ENSample ID listThe following table
- Seite 42 und 43: EN9.4 Selecting the Measurement Mod
- Seite 44 und 45: EN9.5.3 Programming RFID TagsBeside
- Seite 46 und 47: EN3. Hold the allocated or programm
- Seite 48 und 49: EN9.6.2 Filling with the Plastic Sy
- Seite 52 und 53: EN5. Select to export the whole li
- Seite 54 und 55: EN10 Cleaning and Storing the Instr
- Seite 56 und 57: EN11 Maintenance11.1 Changing the B
- Seite 58 und 59: ENFig. 11 - 3Removing the pump lock
- Seite 60 und 61: EN11.3.1 Accessing Device Informati
- Seite 62 und 63: EN12 ReadjustmentTo ensure accurate
- Seite 64 und 65: EN12.5 Restoring to Factory Adjustm
- Seite 66 und 67: EN14 Technical DataMeasuring range:
- Seite 68 und 69: ENAppendix A: Density of Water 4 (0
- Seite 70 und 71: ENID Measuring unit Minimum Maximum
- Seite 72 und 73: ENAppendix C: CE Declarations of Co
- Seite 74 und 75: ENAppendix D: EC-Type-Examination C
- Seite 76 und 77: ENAppendix E: RFID Module Test Repo
- Seite 78 und 79: ENAppendix F: Firmware VersionsFirm
- Seite 80 und 81: ENSetup Measurement Mode Precise; F
- Seite 82 und 83: DEDieses Dokument wurde mit der nö
- Seite 84 und 85: DE9.1.3 Kundenfunktionen importiere
- Seite 86 und 87: DE1 Zur BetriebsanleitungDiese Betr
- Seite 88 und 89: DE2 Sicherheitshinweise• Lesen Si
- Seite 90 und 91: DESchutzmaßnahmen bei leicht entz
- Seite 92 und 93: DE4 DMA 35 - ein ÜberblickDas DMA
- Seite 94 und 95: DETabelle 5.1: Gelieferte TeileSymb
- Seite 96 und 97: DE6 Beschreibung des Gerätes6.1 Vo
- Seite 98 und 99: DE6.2 Rückansicht123465Abb. 6 - 3
9.9 Exporting Measuring Data to a PCENExporting measuring data to a PC does not require a special program. You just needa PC with an infrared interface or a PC with an IrDA USB adapter installed andconnected. Contact your Anton Paar representative if you need an IrDA USB adapterfor your PC.To set the data export format1. Press and select "Setup > Import / Export > Data Format".You can choose between different format options.2. Choose the data format, or , and press .3. Choose the delimiter, [comma], [semi-colon] or [slash], and press.4. Choose the decimal point, [point] or [comma], and press .Your changes are accepted and used for the following data export. The settingsfor delimiter and decimal point are only applied with CSV data exports.To export a measured value1. If your PC has no IrDA interface, install the IrDA USB adapter on your PC.2. Press and select "Measuring Data".3. Select the measured value required for export using the arrow keys.4. Press .5. Hold the IrDA interface of <strong>DMA</strong> <strong>35</strong> to the IrDA interface of the PC to establishcommunication between the two instruments.6. Select to export the selected measured value to the PC.A dialog window appears on the PC asking if you want to accept the file.7. Click .The measured value is transferred and filed on the desktop.To export all measuring dataYou have two options for exporting all measuring data to a PC.Option 1:1. If your PC has no IrDA interface, install the IrDA USB adapter on your PC.2. Press and select "Measuring Data".3. Press .4. Hold the IrDA interface of <strong>DMA</strong> <strong>35</strong> to the IrDA interface of the PC to establishcommunication between the two instruments.C96IB01 51