Test Procedure<br />

Preparation:<br />

Allow test unit and chase buffer to come to room temperature (18 – 28 °C) before opening!<br />

Never store test units at room temperature! Use test units immediately after opening!<br />

Compare batch code on the outerbox label and given batch code at <strong>Milenia</strong> ® POCScan Reader if<br />

necessary enter new batch code and parameters. (see <strong>Milenia</strong> ® POCScan Reader Manual, p. 12/13)<br />

Consult attended documents: Certificat of Analysis (CoA-MQLRP) - batch code specific.<br />

Capillary blood collection:<br />

a) Hold transfer pipet horizontally into the blood drop without pushing the bulb.<br />

The blood is filled in automatically by capillarity-up to the black mark, equivalent to 10 µL.<br />

b) To apply the sample into the sample application port of the test unit squeeze pipet’s bulb to empty<br />

it. Do this in a timely manner but not too fast.<br />

Avoid blood clotting or air bubbles. (Risk of false results)<br />

Steps<br />

1.<br />

Pipet 10 µL whole blood or 5 µL serum into the sample application port<br />

of the test unit.<br />

2.<br />

Add 2 drops chase buffer<br />

Note: Do not contaminat the vial with sample material!<br />

3.<br />

4.<br />

Incubation<br />

10 minutes at room temperature (18 - 28 °C)<br />

Put test unit in the drawer.<br />

Close drawer!<br />

Start<br />

SCAN<br />

- Mode, SCAN IN PROGRESS:<br />

<br />

<br />

Scan<br />

Result calculation<br />

5. Result is displayed.<br />

Scan Mode<br />

CRP: 10 - 20 mg/l<br />

Next test Delete Print<br />

If result is INVALID check described items on flyer:<br />

“Notice INVALID/ A / 2009-10-16<br />

For enquiries or error messages, please contact your local distributor or<br />

Customer Service at <strong>Milenia</strong> <strong>Biotec</strong> <strong>GmbH</strong>.<br />

E mail: cwillsch@milenia-biotec.de<br />

The <strong>Milenia</strong> ® POCScan Reader manual can be downloaded from the<br />

website www.milenia-biotec.de<br />

<strong>Milenia</strong> <strong>Biotec</strong>, Versailler Straße 1, D-35394 Gießen<br />