4 Installation<br />
4.9 Adjusting the UV Calibration<br />
<br />
Calibrating is recommended for initial startup and each time after replacing the lamp.<br />
• Push the MENU button to get to the<br />
main menu.<br />
• The status LED extinguishes.<br />
• Push the UP or DOWN button to get to<br />
the UV CALIBRATION menu option.<br />
• Push SELECT button.<br />
This process is possible not before<br />
5 minutes after reaching the Ready for<br />
Use of the unit because the monitoring<br />
device must have been stabilized.<br />
• Push the UP or DOWN button to select<br />
the desired percentage.<br />
• Push OK button to complete the<br />
operation and accept the setting.<br />
• The status LED lights up.<br />
The main window appears when UV<br />
calibration setup is finished.<br />
UV CALIBRATION<br />
PLEASE WAIT<br />
XXX SEC<br />
23