H 24 REFERENCE GUIDE/ALLGEMEINE INFORMATION<br />

HU-FRIEDY TROUBLE SHOOTING GUIDE<br />

Instruments<br />

Problem Cause Prevention<br />

Spotting • Insufficient rinsing • Rinse thoroughly under steady stream<br />

after ultrasonic of water for 30 seconds<br />

cleaning<br />

• Insufficient drying • Rinse with hot water<br />

after ultrasonic • Optional: Dip cassettes in alcohol<br />

cleaning after rinsing<br />

• Not changing • Solution should be changed at least<br />

ultrasonic solution once a day<br />

• Sterilizer has not • Sterilizers should be cleaned weekly<br />

been cleaned • Use only distilled water for reservoir<br />

Rust • Corrosion from • Separate stainless and carbon<br />

carbon instruments instruments<br />

spreads to stainless • For carbon steel instruments: Dip in<br />

steel instruments pre-sterilized rust-inhibiting solution as<br />

suggested by sterilizer manufacturer<br />

Pitting • Chemical attack • Rinse and dry instruments thoroughly<br />

on instruments • Use approved cleaning, sterilization<br />

solutions only. Never use household bleach<br />

or stain and tartar remover<br />

IMS Cassettes<br />

Problem Cause Prevention<br />

Staining (Resin cassettes only)<br />

Black • Amalgam left in • Thoroughly empty amalgam carrier before returning<br />

carrier to cassette<br />

• If carrier is plugged, sterilize separately and unplug while<br />

carrier is warm<br />

Green • Chrome breakdown • Inspect instruments and replace those with cracked<br />

of instruments handles or peeling plating<br />

Yellow/Brown • Sterilizer has not • Change reservoir water once a week<br />

been cleaned •Regularly clean chamber and filters<br />

• Normal use • Every 6 months soak cassettes (without instruments and<br />

discoloration not in the ultrasonic) in 100% bleach for<br />

5-30 minutes<br />

Broken (Resin and Stainless Steel cassettes)<br />

Hinges • Overloading/ • Instruments should not protrude from cassette<br />

improperly placed • Only light force is needed to close cassette<br />

instruments<br />

• Improperly placed • Do not reposition cassette rails<br />

rails<br />

Wet Packs • Insufficient drying • Thoroughly dry cassettes after cleaning, before wrapping<br />

either before or • Crack open autoclave sterilizer door during dry cycle<br />

during sterilization • Optional: After sterilization cycle, leave cassettes in warm<br />

sterilizer for 10 minutes<br />

• Improper loading • Do not overpack sterilizer<br />

of cassettes in • Keep cassettes slightly separated within the chamber<br />

sterilizer • Always use sterilizer’s cassette rack<br />

25<br />

24<br />

23<br />

22<br />

21<br />

20<br />

19<br />

18<br />

17<br />

16<br />

15<br />

14<br />

13<br />

12<br />

11<br />

10<br />

9<br />

8<br />

7<br />

6<br />

5<br />

4<br />

3<br />

2<br />

1<br />

0<br />

CENTIMETERS<br />

10<br />

9<br />

8<br />

7<br />

6<br />

5<br />

4<br />

3<br />

2<br />

1<br />

0<br />

INCHES

Hurra! Ihre Datei wurde hochgeladen und ist bereit für die Veröffentlichung.

Erfolgreich gespeichert!

Leider ist etwas schief gelaufen!