13.07.2015 Views

Elma BM811 Elma BM817 Elma BM515 - Elma Instruments

Elma BM811 Elma BM817 Elma BM515 - Elma Instruments

Elma BM811 Elma BM817 Elma BM515 - Elma Instruments

SHOW MORE
SHOW LESS
  • No tags were found...

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Elma</strong> 811 + 817 + 515 side 77DC VoltageRANGEAccuracy50.00 mV 0.12% + 2d500.0 mV 0.06% + 2d5.000V, 50.00V, 500.0V, 1000V 0.08% + 2dNMRR : >60dB @ 50/60HzCMRR : >120dB @ DC, 50/60Hz, Rs=1kΩInput Impedance : 10MΩ, 16pF nominal (44pF nominal for 50mV & 500mV ranges)AC VoltageRANGEAccuracy50Hz -- 60Hz50.00mV, 500.0mV, 5.000V, 0.5% + 3d50.00V, 500.0V, 1000V40Hz -- 500Hz50.00mV, 500.0mV0.8% + 3d5.000V, 50.00V, 500.0V1.0% + 4d1000V1.2% + 4dUp to 20kHz50.00mV, 500.0mV0.5dB*5.000V, 50.00V, 500.0V3dB*1000VUnspec'd*Specified from 30% to 100% of rangeCMRR : >60dB @ DC to 60Hz, Rs=1kΩInput Impedance : 10MΩ, 16pF nominal (44pF nominal for 50mV & 500mV ranges)DC CurrentRANGE Accuracy Burden Voltage500.0μA0.15mV/μA5000μA0.15mV/μA50.00mA 0.2%+4d 3.3mV/mA500.0mA3.3mV/mA5.000A 0.03V/A10.00A*0.03V/A*10A continuous, 20A for 30 seconds max with 5 minutes cool down interval